TempStick Probe -80
TempStick Probe -80 temperature data logger is a miniaturized temperature recorder with cable and rigid metal probe of verious lengths, calibrated from …
S-MicroW L Ultra Freeze is a data logger for ultra low temperatures from -80°C to 140°C (calibration from -40°C to 140°C) with 20, 100 mm or on demand length external probe (probes cannot be switched) on a 5 mm base (base height is not counted for probe length), managed with Windows software and USB interface (DiskInterface HS, Multibay). Battery is user replaceable and the data logger is provided with an Accredia (NIST equivalent) traceable cetificate on 6 points.
The data logger looks like the S-MicroW L standard but it includes a calibration from -40°C. With calibration from -40°C it can be used also at -80°C with an accuracy within ± 0,5 °C within range -80 °C ÷ -40 °C. If -80°C is requested to be on the certificate, an extra calibration point for very low temperatures must be purchased. It is particularly indicated for monitoring and validation of ultra freezers.
Note: if used below -30°C, battery life will be highly reduced. Mission max duration at -80°C: 12 hours. Using the logger for long periods below -40°C will constantly reduce mission max duration. For longer missions purchase the -80°C special battery for S-MicroW L Ultra Freeze. With -80°C special battery mission max duration at -80°C with the special battery: 2 months. Using the logger for long periods below -40°C will constantly reduce mission max duration. Max temperature of use of the -80°C battery is 80°C: REMOVE IT AND USE STANDARD BATTERY TO USE THE LOGGER IN AUTOCLVE OR ABOVE 80°C.
The system is made up by:
Dimensions | 39 h X 20 Ø (mm) |
---|---|
Probe dimensions | Probe base dimensions 3 h X 14 Ø (mm) - Probe 20/100/on demand l X 3 Ø (mm) (l on demand: min. 12 mm / max. 175 mm. For longer probes ask for quotation) |
Weight | 50 g |
Materials | Stainless steel AISI316L, PEEK |
Temperature range | -80 °C ÷ +140 °C |
Standard calibration points (temperature) | -40/-20/0/25/50/75/100/121/140°C |
Extra calibration points (temperature) | Within the range -80 °C ÷ +140 °C |
Temperature resolution | 0,01 °C |
Temperature accuracy | ± 0,1 °C (within the calibration range) / ± 0,5 °C (- 80 °C ÷ -40°C without -80°C point) |
Memory (n. of acquisitions) | 20.224 |
Acquisition step | From 1 every second up, with 1 second steps |
Protection degree | IP68 |
Battery life | +10.000.000 acquisitions at 1 second step continuously (calculated time @25°C. Battery life may be shorter if used in low temperatures) |
Software&Mobile App | SPD, TS Manager |
Accessories | DiskInterface HS, Multibay universale |
Alarms | Local (acoustic and/or visual) on PC (for P-Radio, S-Radio only) |