Subscribe to our newsletter - Tech support
1989 / 2024 Flip through our history

Temperature data logger with fast response time up to 250°C provided with Reference certificate

Temperature recording device from 0°C to 250°C (calibration from 25°C to 140°C, above 140°C the probe only resists) with 30 cm bendable probe and 1,5 mm diameter, managed with Windows software and USB interface (DiskInterface HS, Multibay). Battery is user replaceable and the data logger is provided with an Accredia (NIST equivalent) traceable cetificate on 6 points. Extra battery kit included.

Video

Applications

Main features

  • Very fast response time thanks to the probe diameter of only 1.5 mm
  • With bendable cable to place the device near the packages instead on top and with rigid probe for penetration (first and last 30 mm cannot be bent)
  • Completely food grade and waterproof
  • All software calculate lethality value (F0, PU, A0 ecc.)
  • Low battery consumption for an extended battery life
  • User replaceable battery (software shows battery status)
  • Very easy to deploy in any type of package
  • Accredia (NIST equivalent) traceable calibration certificate included
  • Available extended calibration from 0°C to 250°C (order extra calibration points; in case of wide calibration range the accuracy might be worse. Request the possible calibration ranges in advance)

 

Plus

  • Extremely high accuracy and precision: with an accuracy of ± 0,1°C these devices can be employed in any application involving food&beverages, pharmaceuticals, validation, laboratory and medical field
  • High accuracy even outside the calibration range
  • Easy to use even in small and narrow places thanks to the bendable probe
  • Printed reports compliant with health regulations and ISO (data are not editable in the software)

 

The system

The system is made up by:

  • MicroW S Bendable temperature data logger
  • DiskInterface HS or Universal Multibay
  • TS Manager 3 software (compatible with the FDA 21 CFR Part 11 regulation)
Dimensions 14 h X 20 Ø (mm)
Probe dimensions Probe base dimensions 3 h X 14 Ø (mm) - Bendable probe 300/on demand l X 1,5 Ø (mm) (first and last 15 mm cannot be bent)
Weight 17 g
Materials Stainless steel AISI316L, PEEK
Temperature range 0 °C ÷ +140 °C
Sensor temperature range 0 °C ÷ +250 °C
Standard calibration points (temperature) 25/50/75/100/121/140°C
Extra calibration points (temperature) Within the range 0 °C ÷ +250 °C
Temperature resolution 0,01 °C
Temperature accuracy ± 0,1 °C (within the calibration range)
Memory (n. of acquisitions) 20.224
Acquisition step From 1 every second up, with 1 second steps
Protection degree IP68
Battery life 560.000 acquisitions at 1 second step continuously (calculated time @25°C. Battery life may be shorter if used in low temperatures)
Software&Mobile App TS Manager 3, Tecno Calib
Accessories DiskInterface HS, Multibay universale

Documentation