Subscribe to our newsletter - Tech support

Temperature data logger up to 250°C with probe of various lengths provided with an Accredia (NIST equivalent) traceable cetificate

MicroW L is a temperature data logger from -40°C to 140°C with 20 or 50 or 100 or 150 mm or on demand length external probe (probes cannot be switched) on a 3 mm base (base height is not counted for probe length), managed with Windows software and USB interface (DiskInterface HS, Multibay). Battery is user replaceable and the data logger is provided with an Accredia (NIST equivalent) traceable cetificate on 6 points (standard calibration from 25°C to 140°C, calibrable from -40°C to 250°C, only the probe will resist above 140°C. For calibrations above 140°C probe minimum length is 50 mm).

Note: if used below -30°C, battery life will be highly reduced.

Video

Applications

Main features

  • With different lengths rigid probe for penetration
  • Completely food grade and waterproof
  • All software calculate lethality value (F0, PU, A0 ecc.)
  • Low battery consumption for an extended battery life
  • User replaceable battery (software shows battery status)
  • Very easy to deploy in any type of package
  • Accredia (NIST equivalent) traceable calibration certificate included
  • Available extended calibration from -40°C to 250°C (order extra calibration points; in case of wide calibration range the accuracy might be worse. Request the possible calibration ranges in advance)
  • For calibration over 140°C the minimum probe length is 50 mm

Plus

  • Extremely high accuracy and precision: with an accuracy of ± 0,1°C these devices can be employed in any application involving pharmaceuticals, food&beverages, validation, laboratory and medical field
  • High accuracy even outside the calibration range
  • Fast response time thanks to the 3 mm diameter probe
  • Printed reports compliant with health regulations and ISO (data are not editable in the software)

The system

The system is made up by:

  • MicroW L temperature data logger
  • DiskInterface HS or Universal Multibay
  • SPD software or TS Manager software (compatible with the FDA 21 CFR Part 11, Annex 11, GAMP 5 regulations)
Dimensions 39 h X 20 Ø (mm)
Probe dimensions Probe base dimensions 3 h X 14 Ø (mm) - Probe 20/50/100/150/on demand l X 3 Ø (mm) (l on demand: min. 12 mm / max. 175 mm. For longer probes ask for quotation)
Weight 50 g
Materials Stainless steel AISI316L, PEEK
Temperature range -40°C ÷ +140°C
Sensor temperature range -40 °C ÷ +250 °C
Standard calibration points (temperature) 25/50/75/100/121/140°C
Extra calibration points (temperature) Within the range -40 °C ÷ +250 °C
Temperature resolution 0,01 °C
Temperature accuracy ± 0,1 °C (within the calibration range)
Memory (n. of acquisitions) 20.224
Acquisition step From 1 every second up, with 1 second steps
Protection degree IP68
Battery life +10.000.000 acquisitions at 1 second step continuously (calculated time @25°C. Battery life may be shorter if used in low temperatures)
Software&Mobile App SPD, TS Manager
Accessories DiskInterface HS, Multibay universale

Documentation