Subscribe to our newsletter - Tech support

Ultra low temperature data logger from -80°C up to 140°C with probe of various lengths. It is particularly indicated for monitoring and validation of ultra freezers.

MicroW XL Ultra Freeze is an ultra low temperature data logger from -80°C to 140°C (calibration from -40°C to 140°C) with 20, 100 mm or on demand length external probe (probes cannot be switched) on a 5 mm base (base height is not counted for probe length), managed with Windows software and USB interface (DiskInterface HS, Multibay). Battery is user replaceable and the data logger is provided with a Reference traceable cetificate on 6 points.
With calibration from -40°C it can be used also at -80°C with an accuracy within ± 0,5 °C within range -80 °C ÷ -40 °C. If -80°C is requested to be on the certificate, an extra calibration point for very low temperatures must be purchased. It is particularly indicated for monitoring and validation of ultra freezers.

Note: if used below -30°C, battery life will be highly reduced. Mission max duration at -80°C: 12 hours. Using the logger for long periods below -40°C will constantly reduce mission max duration. For longer missions purchase the -80°C special battery for S-MicroW XL Ultra Freeze. With -80°C special battery mission max duration at -80°C with the special battery: more than 2 months. Using the logger for long periods below -40°C will constantly reduce mission max duration. Max temperature of use of the -80°C battery is 80°C: REMOVE IT AND USE STANDARD BATTERY TO USE THE LOGGER IN AUTOCLVE OR ABOVE 80°C.

Video

Applications

Main features

  • With different lengths rigid probe for penetration
  • Completely food grade and waterproof
  • All software calculate lethality value (F0, PU, A0 ecc.)
  • Low battery consumption for an extended battery life
  • User replaceable battery (software shows battery status)
  • Very easy to deploy in any type of package
  • National Reference traceable calibration certificate included from -40°C
  • Available extended calibration from -80°C (order extra calibration points; in case of wide calibration range the accuracy might be worse)

 

Plus

  • Extremely high accuracy and precision: with an accuracy of ± 0,1°C these devices can be employed in any application involving pharmaceuticals, validation, laboratory and medical field
  • High accuracy even outside the calibration range
  • Save money on calibration costs with calibration from -40°C included
  • High accuracy even from -80°C
  • Fast response time thanks to the 3 mm diameter probe
  • Printed reports compliant with health regulations and ISO (data are not editable in the software)

The system

The system is made up by:

  • MicroW XL Ultra Freeze temperature data logger
  • DiskInterface HS or Universal Multibay
  • SPD software or TS Manager software (compatible with the FDA 21 CFR Part 11 regulation)
Dimensions 64 h X 20 Ø (mm)
Probe dimensions Probe base dimensions 3 h X 14 Ø (mm) - Probe 20/100/on demand l X 3 Ø (mm) (l on demand: min. 12 mm / max. 175 mm. For longer probes ask for quotation)
Weight 61 g
Materials Stainless steel AISI316L, PEEK
Temperature range -80 °C ÷ +140 °C
Standard calibration points (temperature) -40/-20/0/25/50/75/100/121/140°C
Extra calibration points (temperature) Within the range -80 °C ÷ +140 °C
Temperature resolution 0,01 °C
Temperature accuracy ± 0,1 °C (within the calibration range)
Memory (n. of acquisitions) 20.224
Acquisition step From 1 every second up, with 1 second steps
Protection degree IP68
Battery life +17.000.000 acquisitions at 1 second step continuously (calculated time @25°C. Battery life may be shorter if used in low temperatures)
Software&Mobile App SPD, TS Manager
Accessories DiskInterface HS, Multibay universale

Documentation

Faq

You may also be interested in...